< 1 s in direct current mode < 1,5 s in pulse current mode < 2 s in alternate current mode
Measurement trigger conditions
Manual or automatic trigger from 2 measures/s to 1 measure/9h, allowing a single operator to be able to perform measurements
EMFs
Automatic compensation of EMF parasites before each measurement for a greater accuracy
Temperature compensation
Choice of metal temperature coefficient Choice of ambient temperature (programmed or measured with external probe) Temperature compensation at 20°C: Resolution: 0.1°C, accuracy: ±0.5°C (R20 = Resistance compensated at ambient temperature equal to 20°C
Temperature coefficient beyond operating range
< 10% accuracy/°C
Relative measurements
Display L = R-R0 or L = 100 x (R-R0)/R0 in % Where L: read value, R: measured value and R: reference value either recalled from memory or entered by the operator
Coil heating calculation
Coil heating calculation according to ambient temperature, original coil resistance at ambient temperature, coil resistance once heated and coil material
Alarms
2 programmable thresholds with visual and sound signal and relay outputs